
Chems Hachani
Founder and CEOSets the evidence strategy with the sponsor and stays accountable for delivery from first scoping call to final report.
A sensor, a pump and an app are certified as separate devices and used as one system. Each needs clinical data for its own claims. Scrutiny of CE-marked glucose sensors is also rising: European diabetologists have proposed minimum accuracy expectations that go further than CE marking requires today.
Eclevar MedTech clinical operations and evaluation team · Updated November 26, 2026 · 9 minute read








Device manufacturers across cardiovascular, orthopedics, wound care and more work with Eclevar. Read all client success stories.
Most diabetes devices that drive treatment decisions are class IIb under the MDR: a CGM used without fingerstick confirmation, an insulin pump, and the algorithm that links them. The EU has no harmonised standard or common specification for CGM accuracy, so each manufacturer has to justify its acceptance criteria against the state of the art, and reviewers increasingly read that state of the art as the published international recommendations.
In practice, the clinical study is designed around three questions. Is the device accurate in every population and glucose range it claims? Does it deliver a measurable benefit, such as time in range? And do the data come from the exact sensor, algorithm and software versions you sell?
The class follows the intended purpose written in the IFU, not the product category. When several rules apply, the strictest one wins.
| Device | Usual class and rule | Evidence the reviewer asks for first |
|---|---|---|
| CGM used for treatment decisions | Class IIb, Rule 10 for monitoring, with the software under Rule 11 | Accuracy across the measuring range and full wear period, in each claimed population and insertion site |
| Insulin pump | Class IIb, Rule 12, as an active device administering a medicinal product | Delivery accuracy and occlusion detection on the bench, then clinical safety in use |
| Automated insulin delivery algorithm | At least the class of the pump it drives, under Annex VIII, section 3.3 | Glycemic outcomes with each CGM and pump combination claimed |
| Connected pen or dose-capture cap | Depends on what the software does, under Rule 11 | Dose-recording accuracy and usability, with more clinical data if the software recommends doses |
Two consequences are easy to miss. A clinical evaluation plan that assumes class IIa for a CGM whose IFU allows insulin dosing from the sensor reading will be challenged before any clinical data are read. And because insulin pumps are class IIb active devices administering a medicinal product, they fall within the clinical evaluation consultation procedure unless an Article 54(2) exemption applies. A prefilled pen sold with its insulin is a medicinal product, and its device part needs a Notified Body opinion under Article 117.
With no EU standard to point to, three sources shape what a Notified Body considers adequate. The FDA's special controls for integrated CGM systems set accuracy criteria that many manufacturers already design to. The IFCC working group on CGM has published recommendations on reporting performance studies (2023) and on comparator data and sampling (2024 and 2025). And in 2024, a consensus of European diabetes specialists funded by IDF Europe proposed minimum expectations for CGMs sold in Europe, which they call eCGM. Our CGM clinical evaluation service builds these expectations into the accuracy evidence.
The eCGM proposal is the most specific of the three. Among other points, it asks for:
These expectations are not law. They are, however, the clearest statement of the current state of the art, and a clinical evaluation that ignores them will have to explain why. A single MARD figure does not answer the question: it hides the direction of the error and changes with the comparator and the glucose distribution of the study.
A clinical evaluation plan we prepared shows a common situation. A European company acted as legal manufacturer for a CGM produced by a manufacturer outside Europe. Under the Directive, equivalence to the producer's own model had been accepted. Under the MDR, that equivalence had to be argued again, and the main accuracy data came from the producer's multicenter study in a non-European population.
| In the file | The reviewer's question | How to close it |
|---|---|---|
| Accuracy study in one non-European population | Does the population match the European intended users, including type 1 share, age and glucose range? | A justified bridging analysis, or a European accuracy study in the populations claimed |
| Equivalence accepted under the Directive | Are the sensor, algorithm and firmware the versions on sale, and is access to the data documented? | A version map, a data access agreement, or a claim built on the device's own data |
| MARD threshold taken from another clinical setting | Do the acceptance criteria fit the intended population and use? | Criteria by glucose range, drawn from current recommendations |
| Clinical benefit left to be defined | What benefit does the user get, and how is it measured? | Benefit endpoints from the time in range consensus, confirmed in PMCF |
| Interfering substances listed in the IFU | Was each one tested, and where is the report? | Interference testing summarized in the CER and aligned with the IFU |
Equivalence to another manufacturer's CGM is possible for a class IIb device, but MDCG 2020-5 requires sufficient access to that manufacturer's data, and an algorithm change alone can break it. Our comparison of equivalence and legacy data routes sets out the tests.
Accuracy shows the device measures well. Benefit shows it helps the person using it. The international consensus on time in range, published in 2019, gives the targets most protocols use for people with type 1 and type 2 diabetes: more than 70 percent of readings between 70 and 180 mg/dL, less than 4 percent below 70 mg/dL and less than 1 percent below 54 mg/dL.
Connected devices make PMCF easier than in most fields, because glucose and insulin data are already recorded. The work is in consent, data protection and linking device data to clinical events in an EDC set up for device data. See how we run PMCF studies under the MDR.
An accuracy study is closer to a laboratory protocol than to a usual device trial. It needs sites that can run in-clinic sessions with frequent sampling and controlled glucose excursions, and a laboratory-grade analyzer on site.
Our clinical investigations team runs these studies from protocol to report, and feeds the results into the clinical evaluation report.
A CGM intended for treatment decisions without fingerstick confirmation is class IIb. The class follows the intended purpose in the IFU, and the strictest applicable rule decides.
No. No harmonised standard or common specification sets CGM accuracy criteria. Manufacturers justify their criteria against the state of the art, including the FDA special controls for integrated CGM systems, the IFCC working group recommendations and the European eCGM consensus.
Yes, if the study meets good clinical practice, used the versions of sensor, algorithm and software on sale, and enrolled a population that matches the European intended users. Otherwise a bridging analysis or a European study is needed.
No. MARD does not show whether errors are high or low, and it changes with the comparator and the glucose distribution of the study. Reviewers expect agreement analyzed by glucose range, wear period and lot.
Article 54 covers class IIb active devices intended to administer a medicinal product, which includes insulin pumps, unless one of the exemptions in Article 54(2) applies.
A small senior team, with European delivery and U.S. clinical consultants for sponsors who plan evidence on both sides of the Atlantic.

Sets the evidence strategy with the sponsor and stays accountable for delivery from first scoping call to final report.

Former clinical lead for final labeling on Medtronic's Hugo robotic-assisted surgery launch team, where he helped design surgeon training and proctoring and started a clinical registry for the platform.
Previously at

Provides strategic input on U.S. medical-device clinical operations, regulatory considerations and real-world evidence programs.

Your first point of contact for scoping, budget scenarios and proposals.
Eclevar leadership includes former Notified Body reviewers. Meet the full team on the leadership page.
Whitepapers, recognition and recordings from our teams and partners.

Whitepaper · BSI and EclevarWritten with Notified Body BSI: a practical reading of the clinical evidence expectations under EU MDR 2017/745, relevant to any high-risk device program.
Read the BSI whitepaper
White paper · TÜV SÜDPublished in the TÜV SÜD knowledge hub with Eclevar's contribution: how breakthrough device technology is approached under the EU MDR.
Read the TÜV SÜD white paper
Top prize in the xShare / EUCROF EHDS & Clinical Research Open Call, presented in Amsterdam. Co-funded by the European Union, Horizon Europe Grant Agreement No. 101136734.
Read the press coverage
Antoine Turzi, CEO of RegenLab, on the randomized PMCF study run across 14 sites in 5 EU countries.
Watch on YouTube
A clinician discusses information capture and patient screening with MILO.
Watch on YouTube
A short introduction to the CER and its clinical evidence.
Watch on YouTubeThis article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The case described is anonymized and reported at the level of the evidence pattern only.
Tell us the device, the intended purpose and populations you claim, and the clinical data you already hold. We return a gap analysis against the MDR and current recommendations, and a study design with sites, sample size, timeline and budget.
We aim to respond within one business day.