Clinical evaluation · Notified Body findings · EU MDR
Article 61(10) lets a manufacturer show conformity without clinical data when that is justified. It does not let the CER skip the clinical benefit. A recent deficiency round on a neonatal resuscitation device shows what a Notified Body asks for, and how to answer without making claims the file cannot support.
Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 6 minute read
Request a review of your 61(10) justification Go to the benefit table01
Article 61(10) allows conformity with the general safety and performance requirements to rest on non-clinical data, if the justification draws on risk management, on how the device interacts with the body and on the performance and claims intended. The CER is still required, and so is a clinical benefit.
The MDR defines a clinical benefit in Article 2(53) as a positive impact on health expressed as a meaningful, measurable, patient-relevant outcome, or a positive impact on patient management or public health. For a device that acts on the patient only through its technical performance, the benefit is indirect. A Notified Body accepts an indirect benefit when each one is tied to a parameter that can be verified on the bench, and to outcomes the literature reports for the device type, without a device-specific outcome claim.
02
The device family was a gas-powered T-piece resuscitator for newborns, delivering air or an air and oxygen mix at set inspiratory and end-expiratory pressures through a circuit and a face mask, with a suction function on some models. The manufacturer had chosen the Article 61(10) route, supported by bench testing to ISO 10651-5 and a state-of-the-art review.
In the second deficiency round, the Notified Body rejected the clinical benefit section because it restated the intended purpose. It asked for an indirect benefit with a meaningful, measurable outcome parameter, and noted that every device has a benefit for the patient, direct or indirect. This is one of the most common findings we see on 61(10) files, together with performance objectives that describe an effect without saying how it is verified.
03
The answer we drafted kept each benefit indirect and gave it two anchors: a bench parameter for the device and an outcome reported in the literature for the device type.
| Indirect benefit | Device parameter verified on the bench | Outcome anchor in the literature |
|---|---|---|
| Effective lung expansion and oxygenation | Accuracy and stability of inspiratory and end-expiratory pressure, to ISO 10651-5 | Controlled pressure delivery with T-piece devices compared with self-inflating bags |
| Less escalation to intubation | The same pressure parameters | Intubation rate reported for the T-piece principle, with no rate claimed for the device itself |
| Correct oxygen delivery | Oxygen concentration accuracy on blender models, by gas analysis | Oxygen targets in neonatal resuscitation guidelines |
| Airway clearance | Suction flow and negative pressure on suction models, to ISO 10079-3 | Guideline recommendations on clearing the airway when it is obstructed |
Intubation rate worked as the outcome parameter because it is recorded in every neonatal resuscitation and marks the line between successful mask ventilation and invasive airway management. We also advised dropping two performance objectives that no standardized bench test supported, since a reviewer would flag them as unverified.
04
For other examples of findings, see what a Notified Body asked on a port CER, and for the PMCF side, how we design PMCF surveys. When a deficiency report has already arrived, our Notified Body deficiency support covers the full response.
Send us the CER, the clinical evaluation plan, the IFU and the deficiency report if you have one. Our clinical evaluation team, which includes former Notified Body reviewers, tells you which benefits and objectives will hold, which need a bench or literature anchor, and what to remove.
Request a review of your 61(10) justification05
It allows conformity with the general safety and performance requirements to be shown without clinical data, when the manufacturer justifies it from risk management, the device's interaction with the body and the intended performance and claims.
Yes. The CER must still state the clinical benefit. For a device that acts through its technical performance, the benefit is indirect, and it should be tied to a measurable parameter.
A benefit to the patient that follows from the device's technical performance rather than from a direct effect, such as better ventilation resulting from accurate and stable airway pressures.
Not a device-specific one without clinical data. It can cite outcomes the literature reports for the device type, as long as the device's own parameters are verified on the bench.
A PMCF plan, or a justification for why PMCF is not needed, is still part of the technical documentation. User surveys and complaint trending are common for this kind of device.
This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The case is reported without the manufacturer, the device names or the Notified Body.