Clinical evaluation · Notified Body findings · EU MDR

Article 61(10) CERs: how to define an indirect clinical benefit a Notified Body accepts

Article 61(10) lets a manufacturer show conformity without clinical data when that is justified. It does not let the CER skip the clinical benefit. A recent deficiency round on a neonatal resuscitation device shows what a Notified Body asks for, and how to answer without making claims the file cannot support.

Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 6 minute read

Request a review of your 61(10) justification Go to the benefit table

01

The short answer

Article 61(10) allows conformity with the general safety and performance requirements to rest on non-clinical data, if the justification draws on risk management, on how the device interacts with the body and on the performance and claims intended. The CER is still required, and so is a clinical benefit.

The MDR defines a clinical benefit in Article 2(53) as a positive impact on health expressed as a meaningful, measurable, patient-relevant outcome, or a positive impact on patient management or public health. For a device that acts on the patient only through its technical performance, the benefit is indirect. A Notified Body accepts an indirect benefit when each one is tied to a parameter that can be verified on the bench, and to outcomes the literature reports for the device type, without a device-specific outcome claim.

02

What the Notified Body asked

The device family was a gas-powered T-piece resuscitator for newborns, delivering air or an air and oxygen mix at set inspiratory and end-expiratory pressures through a circuit and a face mask, with a suction function on some models. The manufacturer had chosen the Article 61(10) route, supported by bench testing to ISO 10651-5 and a state-of-the-art review.

In the second deficiency round, the Notified Body rejected the clinical benefit section because it restated the intended purpose. It asked for an indirect benefit with a meaningful, measurable outcome parameter, and noted that every device has a benefit for the patient, direct or indirect. This is one of the most common findings we see on 61(10) files, together with performance objectives that describe an effect without saying how it is verified.

03

Benefits tied to measurable parameters

The answer we drafted kept each benefit indirect and gave it two anchors: a bench parameter for the device and an outcome reported in the literature for the device type.

Indirect clinical benefits of a neonatal T-piece resuscitator and how each one is verified
Indirect benefitDevice parameter verified on the benchOutcome anchor in the literature
Effective lung expansion and oxygenationAccuracy and stability of inspiratory and end-expiratory pressure, to ISO 10651-5Controlled pressure delivery with T-piece devices compared with self-inflating bags
Less escalation to intubationThe same pressure parametersIntubation rate reported for the T-piece principle, with no rate claimed for the device itself
Correct oxygen deliveryOxygen concentration accuracy on blender models, by gas analysisOxygen targets in neonatal resuscitation guidelines
Airway clearanceSuction flow and negative pressure on suction models, to ISO 10079-3Guideline recommendations on clearing the airway when it is obstructed

Intubation rate worked as the outcome parameter because it is recorded in every neonatal resuscitation and marks the line between successful mask ventilation and invasive airway management. We also advised dropping two performance objectives that no standardized bench test supported, since a reviewer would flag them as unverified.

04

What else had to change in the CER

  • One intended purpose everywhere. The same wording in the CER, the clinical evaluation plan, the IFU and the technical file, with uses the IFU and the testing do not cover removed.
  • An explicit statement. All clinical benefits are indirect and achieved through verified technical performance, in line with Article 61(10).
  • A dedicated justification. One section on risk management results, the limited interaction with the patient, the verified performance and the literature behind each indirect benefit, naming the requirements met through non-clinical evidence.
  • No transitional argument. The justification stands on Article 61(10) alone, without leaning on the transitional provisions.
  • A rewritten benefit-risk section. Built on the new benefits and parameters, and carried into the PMCF plan, which for a family like this usually rests on user surveys and complaint trending.

For other examples of findings, see what a Notified Body asked on a port CER, and for the PMCF side, how we design PMCF surveys. When a deficiency report has already arrived, our Notified Body deficiency support covers the full response.

Have your 61(10) justification read before the next round

Send us the CER, the clinical evaluation plan, the IFU and the deficiency report if you have one. Our clinical evaluation team, which includes former Notified Body reviewers, tells you which benefits and objectives will hold, which need a bench or literature anchor, and what to remove.

Request a review of your 61(10) justification

05

Questions manufacturers ask

What does Article 61(10) of the MDR allow?

It allows conformity with the general safety and performance requirements to be shown without clinical data, when the manufacturer justifies it from risk management, the device's interaction with the body and the intended performance and claims.

Does a device under Article 61(10) still need a clinical benefit?

Yes. The CER must still state the clinical benefit. For a device that acts through its technical performance, the benefit is indirect, and it should be tied to a measurable parameter.

What is an indirect clinical benefit?

A benefit to the patient that follows from the device's technical performance rather than from a direct effect, such as better ventilation resulting from accurate and stable airway pressures.

Can a 61(10) CER claim a clinical outcome for the device?

Not a device-specific one without clinical data. It can cite outcomes the literature reports for the device type, as long as the device's own parameters are verified on the bench.

Is PMCF needed under Article 61(10)?

A PMCF plan, or a justification for why PMCF is not needed, is still part of the technical documentation. User surveys and complaint trending are common for this kind of device.

Sources and scope

  • Regulation (EU) 2017/745, Article 2(53), Article 61(10), Annex I and Annex XIV
  • ISO 10651-5, gas-powered emergency resuscitators
  • ISO 10079-3, suction equipment powered from a vacuum or pressure source
  • International Liaison Committee on Resuscitation, neonatal life support recommendations

This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The case is reported without the manufacturer, the device names or the Notified Body.

Reforming Clinical Evaluation of Medical Devices in Europe