Clinical evaluation · Class III implants · EU MDR
The MDR names surgical meshes as class III implants, whatever their age on the market. Most legacy mesh files have enough patients. What they lack is follow-up long enough to support a recurrence or chronic pain claim, and that is where a Notified Body review usually stops.
Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 8 minute read
Request a mesh CER gap review Go to the gap in legacy files01
Rule 8 of Annex VIII classifies surgical meshes as class III. Under Article 61(4), a class III implant needs a clinical investigation unless an exemption applies. Surgical meshes are not on the list of well-established technologies in Article 61(6)(b), including after the list was widened in June 2026.
A mesh that was lawfully on the market under the Directive can rely on existing clinical data under Article 61(6)(a), but only if those data are sufficient for every claim it keeps. Sufficiency is where most mesh files fail, because the data rarely cover the device lifetime. Legacy class III devices that meet the transition conditions must hold an MDR certificate by December 31, 2027.
02
Most implantable devices are class IIb under Rule 8. Breast implants and surgical meshes are the named exceptions and go to class III. That single line drives several obligations that a class IIb file would not face.
03
A recent clinical evaluation we wrote is a typical case. A family of polyester meshes for groin and ventral hernia, CE marked under the Directive for well over a decade, moving to the MDR as class III. No equivalence was claimed. The clinical data came from the manufacturer's own post-market data collections, the published literature on the device and complaint data.
The data collections covered several hundred patients. They were retrospective, and follow-up ran from about one month to three months. The declared device lifetime was five years, and the main performance claim was a low recurrence rate. A Notified Body reviewer reads those two facts side by side.
| In the legacy file | The reviewer's question | How to close it |
|---|---|---|
| Retrospective series with 1 to 3 months of follow-up | How can short-term data support a recurrence or chronic pain claim over the device lifetime? | Long-term reoperation data from registries now, and a prospective PMCF cohort with clinical and patient-reported follow-up |
| Pooled results across groin and ventral repair | Is each indication, approach and mesh position in the IFU supported by its own data? | Split the analysis by indication, or narrow the IFU to what the data cover |
| Benchmark devices named, no equivalence claimed | Are the acceptance criteria taken from the state of the art, with numbers? | Set quantified performance and safety thresholds from current guidelines and meta-analyses |
| IFU that does not list known mesh complications | Does the labeling reflect the risks the literature reports? | Align IFU, risk file, CER and SSCP on the same list of side effects |
| PMCF plan still to be written | Which residual uncertainty does each PMCF activity answer? | Write the PMCF plan from the CER gaps, one objective per open question |
None of these gaps is about the mesh itself. They are about what the file can prove. MDCG 2020-6 is clear that PMCF can resolve residual questions after certification. It cannot stand in for clinical evidence that is missing at the time of certification.
04
The international HerniaSurge guidelines for groin hernia were published in 2018 and updated in 2023. The update names recurrence and chronic pain as the most frequent problems after groin repair, which makes them the endpoints a reviewer expects first.
Report each endpoint by indication and by approach. Open groin repair, laparoscopic groin repair and intraperitoneal ventral repair carry different risks, and a mesh used in all three needs a benchmark for each. The clinical evaluation report should state the threshold before the data are analyzed, so the conclusion is not read as fitted to the result.
05
Few device fields have registries as mature as hernia surgery. The Swedish Hernia Register started in 1992, the Danish Hernia Database in 1998 and extended to ventral hernia in 2007, and Herniamed, covering Germany, Austria and Switzerland, in 2009. EuraHS records ventral and incisional repair across Europe, and Club Hernie covers France.
Registry data can give a mesh file what its own studies lack: thousands of patients and years of reoperation data. Before building a PMCF plan on a registry, check four points.
The stronger plan usually combines both: a registry analysis for long-term reoperation, and a prospective cohort for clinical recurrence, pain and patient-reported outcomes. See how we design PMCF studies under the MDR and registry-based evidence.
06
For a legacy class III mesh under transition, the MDR certificate is due by December 31, 2027. A prospective study started now cannot produce two-year data by then. The file submitted for certification has to stand on the evidence that exists today, with a PMCF plan that a reviewer accepts as answering the questions that remain.
The proposed MDR revision, COM(2025)1023, is not law. Plan on the regulation as it stands. If the mesh was CE marked through an equivalence claim, read our comparison of equivalence and legacy data routes first.
Send us your CER, PMCF plan and IFU. Our clinical evaluation team, which includes former Notified Body reviewers, marks each claim against the data that supports it and tells you what can be closed before 2027 and what needs PMCF.
Request a mesh CER gap review07
Yes. Rule 8 of Annex VIII classifies surgical meshes, like breast implants, as class III implantable devices.
No. Delegated Regulation (EU) 2026/1451 widened the list in Article 61(6)(b), but surgical meshes are not on it. A clinical investigation remains the default under Article 61(4) unless another exemption applies.
It can under Article 61(6)(a), if it was lawfully placed on the market under the Directive and its clinical evaluation rests on sufficient clinical data. The Notified Body judges sufficiency, claim by claim, against the state of the art.
Yes, as one PMCF activity, provided the registry identifies the device at product level, captures the outcomes you need and gives you lawful access under an agreed analysis plan.
Long enough to cover the declared lifetime and the endpoints claimed. Recurrence and chronic pain develop over years, so a follow-up of a few months cannot support those claims. The PMCF plan should justify the duration against both.
This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The case described is anonymized and reported at the level of the evidence pattern only.