EU MDR clinical evaluation · Legacy devices

Equivalence or legacy data under the EU MDR: choosing the route claim by claim

A device moving from the Directive to the MDR can lean on a comparator or on its own history. MDCG 2020-5 and MDCG 2020-6 set different conditions for each, and files stall when the two are mixed. This guide compares the routes, then covers the two decisions that usually decide which one holds: classification and the literature method.

Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 9 minute read

Request a CER gap review proposal Go to the equivalence or legacy comparison

01

The short answer

A manufacturer moving a device from the Medical Devices Directive to the EU MDR has two ways to use existing evidence in its clinical evaluation report (CER). It can claim equivalence to a named device, or it can rely on the clinical data its own legacy device generated under the Directive. Equivalence needs one comparator that matches on technical, biological and clinical characteristics. Legacy data needs enough clinical data on the device itself for every claim in the instructions for use. When neither holds for a claim, the choice is to generate new clinical data or to remove the claim.

The decision should be made claim by claim, before the CER is written, and recorded in the clinical evaluation plan.

02

Where legacy files fall short

A file that satisfied the Directive was rarely built to show clinical benefit for each indication. The EU MDR asks for exactly that: a clinical evaluation under Article 61 and Annex XIV that is planned, device specific and kept up to date for as long as the device is sold.

Three gaps appear in most legacy files we review:

  • Claims without their own data. A product family lists several functions or indications, but clinical data exist for only some of them.
  • Evidence that is about the technology, not the device. Bench results, standards compliance and data on similar products are strong, while clinical data on the device itself are thin.
  • Post-market data that were never collected with a question in mind. Complaint and sales figures exist, but no proactive follow-up that a reviewer can count as clinical evidence.

None of these blocks an MDR file. Each one changes which route is realistic.

03

Equivalence or legacy data: how to choose

The two routes rest on different guidance. Equivalence follows MDCG 2020-5. Legacy device evidence follows MDCG 2020-6. They are often presented as interchangeable, and they are not.

Equivalence and legacy data compared for a device previously CE marked under the Directive
QuestionEquivalenceLegacy device data
What it relies onClinical data from one named device shown to be equivalentClinical data generated on your own device before and after CE marking
Can the comparator hold a Directive certificate?Yes for devices that are neither implantable nor class III. For implantable and class III devices the comparator must be MDR certifiedNot applicable. No comparator is named
Role of similar devicesNone. Parts of several devices cannot be combined into one claimSupporting information only, and only for well-established technologies
MaterialsSame materials in contact with the same tissues. The older exceptions no longer applyNot a test, since the data are your own
Where it breaksOne claim or function for which the comparator has no dataAn indication with no clinical data on the device
Role of PMCFExpected alongside the claim, through PMCF or proactive post-market surveillanceConfirms conclusions. It should not be the plan for filling a gap after certification

When equivalence to your own predecessor works

For a class IIa or non-implantable class IIb device, a manufacturer can claim equivalence to its own earlier model even if that model was only ever certified under the Directive. Access to the data is not an issue because both files are yours. The comparator still has to match on every characteristic, and it has to carry clinical data for every claim you keep.

A pattern we see often: the new model adds one measurement mode that the predecessor never had. The equivalence table looks complete until a reviewer asks for the data behind that mode. There are then three options. Group the endpoints so the claim is assessed as part of a wider function, generate data, or remove the claim from the instructions for use and the marketing material. The first is the weakest in front of a Notified Body.

When legacy data is the stronger base

If the device has years of European use and the manufacturer can extract clinical data from that use, the legacy approach avoids arguing equivalence at all. The guidance treats Directive-era clinical data as part of the evidence at the first MDR conformity assessment. It also says that data which were sufficient under the Directive are not automatically sufficient under the MDR.

The limit is easy to miss. Data from similar devices are not clinical data on your device. For a device that is not a well-established technology, they cannot carry a claim.

Plan for the next certificate, not only this one

A comparator that left the market years ago becomes harder to defend at each renewal. Whichever route supports the first MDR certificate, the practical advice is the same: start a PMCF program that produces clinical data on the current device, so the next clinical evaluation stands on its own data. See how we design PMCF studies under the MDR.

04

Classification and the word implantable

The MDR definition of an implantable device turns on two things: the device is introduced by clinical intervention, and it is intended to stay in place afterward, for at least 30 days when only partly introduced. Whether it touches tissue directly is not part of the definition.

This catches product types that their manufacturers have never thought of as implants. Dental materials are the common example. A bonding agent or primer used to fix a crown may be assessed on its worst-case use, and a material supplied to make a custom-made restoration may be judged by whether the finished restoration is implantable. The same product can therefore sit in a different Notified Body code, and face a different clinical evidence expectation, than the manufacturer planned for.

The consequence is practical. Implantable status brings a summary of safety and clinical performance, tighter technical documentation review and, unless an exemption applies, a clinical investigation requirement. Settle the classification argument with your Notified Body in writing before the clinical evaluation plan is finalized. Our SSCP page explains what that document must contain.

05

A literature search that survives review

Reviewers rarely reject a CER because of what the literature says. They reject it because they cannot reproduce how the literature was found. For catheters, guidewires and other accessory devices, where the published evidence is mostly about the procedure, the method carries the file.

A search protocol that holds up has five parts, written before any search is run:

  • Separate searches for the state of the art, for named benchmark devices and for the device under evaluation.
  • Questions framed by population, intervention, comparator and outcome, with the exact search strings and date limits.
  • Safety and performance endpoints defined as measurable rates, so the state of the art yields acceptance thresholds.
  • Appraisal criteria for suitability and data contribution, with coded exclusion reasons.
  • Named vigilance databases to be screened, in Europe and the United States.

Japanese-language publications and domestic post-market data can be used. The CER has to justify why results in a Japanese population transfer to European practice.

06

Deadlines and the 2026 changes

For legacy devices that meet the transition conditions, the period ends on December 31, 2027 for class III and class IIb implantable devices, and on December 31, 2028 for other class IIb, class IIa and class I sterile or measuring devices.

Two delegated regulations published on June 29, 2026 widened the lists of well-established technologies. Regulation (EU) 2026/1451 adds device types to Article 61(6)(b), including guidewires, several catheter types, dental implants, orthodontic devices, dental barriers and dental veneers. For those types a clinical investigation is not mandatory, provided the clinical evaluation rests on sufficient clinical data. It does not remove the CER or PMCF. Our guide to the 2026 WET list covers which device types qualify. Surgical meshes are not among them, as our article on hernia and surgical mesh under the MDR explains.

The wider revision of the MDR proposed by the European Commission in December 2025 is still in negotiation between the European Parliament and the Council. It is not law, and files submitted now are assessed under the current text.

07

For manufacturers based in Japan

The equivalence question is sharper for devices first approved in Japan, because many reached that market on substantial equivalence to a predicate, with no clinical data on the device itself. Our Japan to EU guide covers the full route, from the authorized representative to EUDAMED.

Decisions on claims and indications are usually taken at headquarters in Japan, while the Notified Body reviewer sits in Europe and reads the file in English. Delay comes from the distance between the two: a question arrives, is translated, discussed and answered weeks later with wording the reviewer did not ask for.

Eclevar works with Japanese manufacturers through Eclevar Japan K.K. in Tokyo and its European teams. Our clinical evaluation staff include former Notified Body reviewers. We also provide EU authorized representative and PRRC services for manufacturers with no legal entity in the Union, and the full clinical evaluation report service when the gap review shows the CER needs rebuilding.

Have your route checked before the Notified Body does

A CER gap review reads your current clinical evaluation against Annex XIV and the two guidance documents above. You receive a written report that states, claim by claim, whether equivalence, legacy data or new evidence is the defensible route, and what PMCF has to cover. The scope is fixed, and the proposal follows a scoping call.

Request a CER gap review proposal

08

Questions manufacturers ask

Can equivalence be claimed to more than one device?

Yes. Under MDCG 2020-5 each presumed equivalent device must match the device under evaluation on all technical, biological and clinical characteristics, and each must be documented in the CER. Characteristics of different devices cannot be combined to build one equivalence claim.

Can we claim equivalence to our own device that was certified only under the MDD?

For devices that are neither implantable nor class III, MDCG 2020-5 allows equivalence to a device certified under the Directives. For implantable and class III devices, the equivalent device must be certified under the MDR. In both cases the comparator must match on technical, biological and clinical characteristics.

Are data from similar devices enough for a legacy device?

No. Under MDCG 2020-6, data from similar devices are supporting information, and they can support conformity only for well-established technologies. Other legacy devices need clinical data on the device itself or on a demonstrated equivalent.

Do the June 2026 delegated regulations remove the need for clinical data?

No. For the device types added to Article 61(6)(b), a clinical investigation is no longer mandatory when existing clinical data are sufficient. A clinical evaluation plan, a CER based on sufficient clinical data, PMCF and, where applicable, an SSCP remain required.

What are the EU MDR transition deadlines for legacy devices?

December 31, 2027 for class III and class IIb implantable legacy devices, and December 31, 2028 for other class IIb, class IIa and class I sterile or measuring legacy devices, provided the transition conditions are met.

Sources and scope

This article is general information for regulatory and clinical affairs teams. It is not a regulatory opinion on a specific device, and it does not predict the outcome of any conformity assessment. Situations described are composites drawn from clinical evaluation work and do not identify any manufacturer.