Cardiovascular · Clinical evaluation · EU MDR
A guidewire that touches the coronary or cerebral arteries is a class III device. Since June 2026, guidewires and pressure wires are also on the list of well-established technologies that can reach CE marking without a new clinical investigation. The exemption is real, and it moves the whole question to one word: sufficient.
Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 7 minute read
Request a guidewire evidence review Go to the gap in imported files01
Coronary and neurovascular guidewires are class III. Delegated Regulation (EU) 2026/1451 added guidewires and pressure wires to Article 61(6)(b) of the MDR, so a clinical investigation is no longer mandatory when the clinical evaluation rests on sufficient clinical data. The CER, the PMCF plan and the summary of safety and clinical performance are still required.
The change matters most to a manufacturer that sells a wire in Japan, the United States or Asia and wants to enter Europe. It only works if the data from those markets and from the literature are on the same model, in the indications claimed, and analyzed well enough for a Notified Body to call them sufficient.
02
Under Rules 6 and 7 of Annex VIII, a surgically invasive device intended for direct contact with the heart or the central circulatory system is class III. The MDR's definition of the central circulatory system includes the coronary arteries, the carotid arteries and the cerebral arteries, so coronary and neurovascular wires sit in class III. The same definition puts aortic grafts in class III (see vascular grafts under the MDR). A wire used only in peripheral vessels is usually class IIa.
Whether Rule 6 or Rule 7 applies depends on the duration of use the manufacturer justifies. Most procedures stay under the 60-minute limit for transient use, but long chronic total occlusion procedures can exceed it. Either way the class is the same. A guidewire is not implantable, so it does not go through the clinical evaluation consultation procedure and needs no implant card.
03
| Requirement | Before June 2026 | After June 2026 |
|---|---|---|
| Clinical investigation | The default under Article 61(4), unless equivalence or legacy data applied | Not mandatory if the clinical evaluation rests on sufficient clinical data |
| Clinical evaluation report | Required | Required, with the sufficiency argument made explicitly |
| PMCF | Required | Required. Any reduction in PMCF activity has to be justified |
| Summary of safety and clinical performance | Required for class III | Required for class III |
The text of Article 61(6)(b) is not limited to devices already sold in Europe. A wire new to the EU can use it, but its evidence has to come from somewhere: use outside Europe, the published literature on the model, and complaint and vigilance data. MDCG 2020-6 accepts that for well-established technologies, sources lower in its evidence hierarchy can carry more weight when they are robust. Agree the route with your Notified Body early. Our guide to the 2026 WET list covers the other device types added.
04
A clinical evaluation plan we wrote for a Japanese manufacturer shows the typical case. A family of coronary guidewires for percutaneous coronary intervention had been sold for several years in Japan, the United States and Asia, with tens of thousands of units, but had never been CE marked. The plan relied on post-market data from those markets and on the literature.
| In the file | The reviewer's question | How to close it |
|---|---|---|
| Units sold and a complaint count | What is the rate of each serious event, and how was it coded? | Rates per 1,000 units by event type, with searches of the FDA and Japanese vigilance databases |
| Literature on the wire family | Is each paper on this model and tip variant, in the indication claimed? | A map of every publication to model, tip and indication, such as chronic total occlusion |
| A material or supplier change since the data were collected | Does the data from before the change still apply? | Bench and biological evaluation results summarized in the CER |
| Use outside Europe | Does practice there match European practice for these lesions? | A justification of applicability, and European user feedback in PMCF |
| No clinical investigation planned | On what legal basis, and is the data sufficient for every claim? | The Article 61(6)(b) route, with a claim-by-claim sufficiency analysis |
If the wire is a new generation of one already certified, equivalence to the earlier model can add data, with full access to its file and an MDR-certified comparator for a class III device. Our comparison of equivalence and legacy data routes sets out the tests, and our guide for Japanese manufacturers covers the wider route to Europe.
05
Coating separation deserves its own line. The FDA issued a safety communication on lubricious coating separation from intravascular devices in 2015 and published labeling and performance testing guidance for guidewires in 2019. ISO 11070 covers the bench side. A European reviewer will expect the CER and the risk file to address the same issue.
For PMCF, a structured survey of interventional cardiologists, literature monitoring and complaint trending often fit a well-established wire better than a new study. See how we run PMCF surveys and PMCF studies under the MDR.
Send us the device description, the IFU and the data you hold from other markets. Our clinical evaluation team, which includes former Notified Body reviewers, tells you whether the Article 61(6)(b) route holds for each claim and what is missing, in English or Japanese.
Request a guidewire evidence review06
Class III. Rules 6 and 7 of Annex VIII place surgically invasive devices in direct contact with the heart or central circulatory system in class III, and that system includes the coronary and cerebral arteries.
Yes. Delegated Regulation (EU) 2026/1451, published in June 2026, added guidewires and pressure wires to Article 61(6)(b).
Not if its clinical evaluation rests on sufficient clinical data. Where the data are not sufficient for a claim, a clinical investigation or a narrower claim is still needed.
Yes, if the data are on the same model and variants, in the indications claimed, and analyzed with event rates and a justification that clinical practice is comparable to Europe.
No. PMCF remains required. Its activities can be proportionate to a well-established device, such as surveys and literature monitoring, but any reduction has to be justified.
This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The case described is anonymized and reported at the level of the evidence pattern only.