Vascular · Clinical evaluation · EU MDR
Whether a vascular graft is class IIb or class III depends on where it sits and what it is coated with, and that decides how much clinical data the file needs. Two recent projects, a CER for collagen-coated aortic grafts and a study outline for a peripheral hybrid graft, show where vascular graft files are usually challenged.
Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 7 minute read
Request a vascular graft evidence review Go to the common gaps01
Every vascular graft is an implantable device, so Article 61(4) of the MDR makes a clinical investigation the default. A graft in contact with the aorta, or one coated with collagen, gelatin or heparin, is class III. A plain synthetic graft for a peripheral artery is class IIb.
The default can be set aside only on a defined route: a legacy device lawfully placed on the market under the Directive with sufficient clinical data, or equivalence to a device you have full access to through a contract. For a class III implant, the clinical evaluation also goes through the expert panel consultation procedure unless an exemption applies.
02
| Graft | Class | Why |
|---|---|---|
| Thoracic or abdominal aortic graft, any material | Class III | Rule 8: direct contact with the central circulatory system, which includes the aorta down to its bifurcation |
| Polyester graft sealed with collagen or gelatin | Class III | Rule 18: a derivative of animal origin, whatever the position |
| Heparin-bonded peripheral graft | Class III | Rule 14: an ancillary medicinal substance, with consultation of a medicines authority |
| Uncoated ePTFE or polyester graft below the aortic bifurcation | Class IIb | Rule 8: long-term implantable device outside the central circulatory system |
All of them need an implant card under Article 18, since vascular grafts are not among the exempted implants. The coronary and cerebral arteries are also part of the central circulatory system, as our article on coronary guidewires explains for wires.
03
A CER we prepared for a manufacturer of collagen-coated woven polyester grafts for the ascending aorta, the arch and the descending aorta, and a study outline for a hybrid graft combining a prosthetic conduit with a stent segment for femoropopliteal bypass, show the gaps that come up most often.
| In the file | The reviewer's question | How to close it |
|---|---|---|
| One CER for aortic and peripheral grafts | Which data support each position, diameter and indication? | Evidence mapped per position and indication, with separate benefit-risk conclusions where the populations differ |
| Benchmarks from a different territory | Are aortic grafts a valid benchmark for a femoropopliteal bypass? | Benchmarks from the same territory and indication: vein and prosthetic bypass, stents and drug-coated balloons for the peripheral claim |
| Literature on polyester grafts in general | Do the data cover this weave, this coating and this porosity? | Device-specific data, or an equivalence argument that addresses the coating and its biological effect |
| A change of coating or supplier | Do earlier data still apply? | Bench data to ISO 7198, biocompatibility, and post-change data from PMS and PMCF |
| A new hybrid design | Is there any clinical data on the combination? | A pre-market clinical investigation sized on patency, with equivalence used only for the component that truly matches |
When equivalence is the plan, our comparison of equivalence and legacy data routes sets out what a reviewer needs for each claim.
04
The state of the art for vascular grafts is set by the vascular surgery guidelines, the reporting standards for lower limb ischemia and the bench requirements of ISO 7198. A reviewer expects the CER and any study to use the same definitions.
Our vascular device CRO team runs the investigations, and our CER team, which includes former Notified Body reviewers, writes and challenges the clinical evaluation.
Send us the product list with positions, coatings and diameters, the IFU and the current CER. We tell you the class of each graft, whether a clinical investigation can be avoided, which benchmarks and endpoints fit each claim, and what PMCF is proportionate.
Request a vascular graft evidence review05
Class III if it is in contact with the aorta or another part of the central circulatory system, or if it contains collagen, gelatin or heparin. An uncoated synthetic graft for a peripheral artery is class IIb under Rule 8.
By default, yes, because it is implantable. Article 61 allows exceptions for legacy devices with sufficient clinical data and for equivalence to a device the manufacturer has full access to.
Collagen and gelatin are derivatives of animal origin, and Rule 18 places devices made with non-viable animal tissues or derivatives in class III.
Primary, assisted primary and secondary patency at 12 months, major amputation, amputation-free survival, reintervention, mortality and device failure.
Yes. Article 18 exempts only a short list of implants such as sutures, staples, screws and plates, and vascular grafts are not on it.
This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The projects described are reported without the manufacturer, the devices or the benchmark products.