Dental · PMCF · EU MDR

Dental implants after the 2026 WET list: when a PMCF study is still worth running

Since June 2026, dental implants are on the MDR list of well-established technologies, so a new clinical investigation is no longer the default. PMCF has not gone away, and most implant files still lack the one thing a reviewer looks for first: data for each implant range and connection the IFU claims.

Eclevar MedTech clinical operations team · Updated November 26, 2026 · 7 minute read

Request a dental PMCF design Go to the study design

01

The short answer

Dental implants are long-term implantable devices, class IIb under Rule 8 of the MDR. Delegated Regulation (EU) 2026/1451 added them to Article 61(6)(b), so a clinical investigation is not mandatory when the clinical evaluation rests on sufficient clinical data. The CER and a PMCF plan are still required.

A prospective PMCF study remains the right tool when the literature does not cover your own implant ranges, connections or prosthetic options, which is the usual case for a mid-sized manufacturer. Designed well, it runs as an observational study within routine care and costs far less than a clinical investigation.

02

What the WET listing changes, and what it does not

  • Changes. A clinical investigation under Article 61(4) is no longer the default for dental implants if existing data are sufficient for every claim.
  • Stays. The clinical evaluation report, the PMCF plan and its evaluation report. Any reduction in PMCF activity has to be justified.
  • Still decides. Sufficiency, claim by claim. Published data on implants in general do not show the performance of your connection, your surface or your narrow-diameter range.

Our guide to the 2026 WET list covers the other device types added, and our dental device CRO services cover the wider dental portfolio.

03

Where implant files fall short

Typical gaps in a dental implant clinical evaluation, and how to close them
In the fileThe reviewer's questionHow to close it
Literature on implants with similar designsWhich data are on this implant, this surface and this connection?Device-specific data for each range, from PMCF or from published studies on the device itself
Several ranges and prosthetic options in one IFUIs each range, diameter and prosthetic option supported?A minimum number of implants per range, and prosthetic type recorded at loading
Survival claims over many yearsDoes the follow-up match the claim?Claims sized to the follow-up available, with long-term data from the literature named
Complications counted per patientAre patients with several implants handled correctly?Analysis per implant with clustering by patient set in the statistical analysis plan

04

A PMCF study that answers the gaps

A PMCF study we manage for a European implant manufacturer shows a design that fits most portfolios. It is a prospective, multicenter, observational study within routine care, covering several implant ranges with different connections and three prosthetic options, with 18 months of follow-up after surgery.

  • Primary endpoint. Osseointegration at loading, between two and six months after surgery, compared with a reference rate using a one-sided confidence interval.
  • Secondary endpoints. Implant stability quotient, marginal bone loss on radiographs, maintained osseointegration, biological and mechanical complications, implant survival and patient satisfaction.
  • Visits. Pre-operative, surgery, loading, 12 months and 18 months, all matching routine implant care, with data captured in an eCRF.
  • Sample size. Counted in implants rather than patients, since patients often receive more than one, with a minimum number of implants per range so that each range can be reported.
  • Sites. Experienced implant surgeons in hospital and private practice, trained on the protocol and good clinical practice before the first patient.

The study answers the device-specific gaps within two years. Long-term survival claims still rest on the literature and on continued PMCF. Our biostatistics team handles the per-implant analysis, and our PMCF team runs the study.

05

Running it in France

France is a practical country for dental PMCF: a dense network of implant surgeons in private practice and hospitals, and a clear route for observational studies. When the study stays within the CE-marked purpose and adds no burdensome procedure, it needs a favorable opinion from an ethics committee (CPP), information to the ANSM and a CNIL reference methodology for the data. If the protocol adds invasive or burdensome procedures, it becomes a PMCF investigation notified under Article 74(1) of the MDR. See our guide to submitting a device study in France.

Get a PMCF design sized to your implant ranges

Tell us your implant ranges, connections and prosthetic options, and the data you already hold. We return a PMCF design with endpoints, sample size per range, countries, sites, timeline and budget, and tell you where existing data already suffice.

Request a dental PMCF design

06

Questions manufacturers ask

What class are dental implants under the EU MDR?

Class IIb. They are long-term implantable devices under Rule 8 of Annex VIII, and they are placed in bone rather than in the teeth.

Are dental implants on the MDR well-established technology list?

Yes. Delegated Regulation (EU) 2026/1451 added dental implants to Article 61(6)(b) in June 2026.

Do dental implants still need PMCF after the WET listing?

Yes. The listing removes the default clinical investigation, not PMCF. Any reduction in PMCF activity has to be justified in the plan.

What is a typical primary endpoint for a dental implant PMCF study?

Osseointegration at loading, usually two to six months after surgery, with implant survival, marginal bone loss and complications as secondary endpoints.

Should a dental implant study count patients or implants?

Implants, with the analysis accounting for patients who receive several. The sample size should also set a minimum number of implants for each range claimed.

Sources and scope

This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The study described is reported without the manufacturer, the devices or the sites.