Since June 2026, dental implants are on the MDR list of well-established technologies, so a new clinical investigation is no longer the default. PMCF has not gone away, and most implant files still lack the one thing a reviewer looks for first: data for each implant range and connection the IFU claims.
Eclevar MedTech clinical operations team · Updated November 26, 2026 · 7 minute read
Request a dental PMCF design Go to the study design01
Dental implants are long-term implantable devices, class IIb under Rule 8 of the MDR. Delegated Regulation (EU) 2026/1451 added them to Article 61(6)(b), so a clinical investigation is not mandatory when the clinical evaluation rests on sufficient clinical data. The CER and a PMCF plan are still required.
A prospective PMCF study remains the right tool when the literature does not cover your own implant ranges, connections or prosthetic options, which is the usual case for a mid-sized manufacturer. Designed well, it runs as an observational study within routine care and costs far less than a clinical investigation.
02
Our guide to the 2026 WET list covers the other device types added, and our dental device CRO services cover the wider dental portfolio.
03
| In the file | The reviewer's question | How to close it |
|---|---|---|
| Literature on implants with similar designs | Which data are on this implant, this surface and this connection? | Device-specific data for each range, from PMCF or from published studies on the device itself |
| Several ranges and prosthetic options in one IFU | Is each range, diameter and prosthetic option supported? | A minimum number of implants per range, and prosthetic type recorded at loading |
| Survival claims over many years | Does the follow-up match the claim? | Claims sized to the follow-up available, with long-term data from the literature named |
| Complications counted per patient | Are patients with several implants handled correctly? | Analysis per implant with clustering by patient set in the statistical analysis plan |
04
A PMCF study we manage for a European implant manufacturer shows a design that fits most portfolios. It is a prospective, multicenter, observational study within routine care, covering several implant ranges with different connections and three prosthetic options, with 18 months of follow-up after surgery.
The study answers the device-specific gaps within two years. Long-term survival claims still rest on the literature and on continued PMCF. Our biostatistics team handles the per-implant analysis, and our PMCF team runs the study.
05
France is a practical country for dental PMCF: a dense network of implant surgeons in private practice and hospitals, and a clear route for observational studies. When the study stays within the CE-marked purpose and adds no burdensome procedure, it needs a favorable opinion from an ethics committee (CPP), information to the ANSM and a CNIL reference methodology for the data. If the protocol adds invasive or burdensome procedures, it becomes a PMCF investigation notified under Article 74(1) of the MDR. See our guide to submitting a device study in France.
Tell us your implant ranges, connections and prosthetic options, and the data you already hold. We return a PMCF design with endpoints, sample size per range, countries, sites, timeline and budget, and tell you where existing data already suffice.
Request a dental PMCF design06
Class IIb. They are long-term implantable devices under Rule 8 of Annex VIII, and they are placed in bone rather than in the teeth.
Yes. Delegated Regulation (EU) 2026/1451 added dental implants to Article 61(6)(b) in June 2026.
Yes. The listing removes the default clinical investigation, not PMCF. Any reduction in PMCF activity has to be justified in the plan.
Osseointegration at loading, usually two to six months after surgery, with implant survival, marginal bone loss and complications as secondary endpoints.
Implants, with the analysis accounting for patients who receive several. The sample size should also set a minimum number of implants for each range claimed.
This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The study described is reported without the manufacturer, the devices or the sites.