Vascular · Clinical evaluation · EU MDR

Vascular grafts under the EU MDR: class, clinical investigation and the endpoints a reviewer expects

Whether a vascular graft is class IIb or class III depends on where it sits and what it is coated with, and that decides how much clinical data the file needs. Two recent projects, a CER for collagen-coated aortic grafts and a study outline for a peripheral hybrid graft, show where vascular graft files are usually challenged.

Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 7 minute read

Request a vascular graft evidence review Go to the common gaps

01

The short answer

Every vascular graft is an implantable device, so Article 61(4) of the MDR makes a clinical investigation the default. A graft in contact with the aorta, or one coated with collagen, gelatin or heparin, is class III. A plain synthetic graft for a peripheral artery is class IIb.

The default can be set aside only on a defined route: a legacy device lawfully placed on the market under the Directive with sufficient clinical data, or equivalence to a device you have full access to through a contract. For a class III implant, the clinical evaluation also goes through the expert panel consultation procedure unless an exemption applies.

02

Classification by position and coating

How common vascular grafts are classified under Annex VIII of the MDR
GraftClassWhy
Thoracic or abdominal aortic graft, any materialClass IIIRule 8: direct contact with the central circulatory system, which includes the aorta down to its bifurcation
Polyester graft sealed with collagen or gelatinClass IIIRule 18: a derivative of animal origin, whatever the position
Heparin-bonded peripheral graftClass IIIRule 14: an ancillary medicinal substance, with consultation of a medicines authority
Uncoated ePTFE or polyester graft below the aortic bifurcationClass IIbRule 8: long-term implantable device outside the central circulatory system

All of them need an implant card under Article 18, since vascular grafts are not among the exempted implants. The coronary and cerebral arteries are also part of the central circulatory system, as our article on coronary guidewires explains for wires.

03

Where graft files are challenged

A CER we prepared for a manufacturer of collagen-coated woven polyester grafts for the ascending aorta, the arch and the descending aorta, and a study outline for a hybrid graft combining a prosthetic conduit with a stent segment for femoropopliteal bypass, show the gaps that come up most often.

Typical gaps in a vascular graft clinical evaluation, and how to close them
In the fileThe reviewer's questionHow to close it
One CER for aortic and peripheral graftsWhich data support each position, diameter and indication?Evidence mapped per position and indication, with separate benefit-risk conclusions where the populations differ
Benchmarks from a different territoryAre aortic grafts a valid benchmark for a femoropopliteal bypass?Benchmarks from the same territory and indication: vein and prosthetic bypass, stents and drug-coated balloons for the peripheral claim
Literature on polyester grafts in generalDo the data cover this weave, this coating and this porosity?Device-specific data, or an equivalence argument that addresses the coating and its biological effect
A change of coating or supplierDo earlier data still apply?Bench data to ISO 7198, biocompatibility, and post-change data from PMS and PMCF
A new hybrid designIs there any clinical data on the combination?A pre-market clinical investigation sized on patency, with equivalence used only for the component that truly matches

When equivalence is the plan, our comparison of equivalence and legacy data routes sets out what a reviewer needs for each claim.

04

Endpoints and benchmarks

The state of the art for vascular grafts is set by the vascular surgery guidelines, the reporting standards for lower limb ischemia and the bench requirements of ISO 7198. A reviewer expects the CER and any study to use the same definitions.

  • Peripheral bypass. Primary, assisted primary and secondary patency at 12 months as defined in the reporting standards, major amputation, amputation-free survival, reintervention, mortality and device failure.
  • Aortic replacement. Graft-related reintervention, bleeding through the graft wall, graft infection, anastomotic false aneurysm, graft dilatation over time and all-cause mortality.
  • Population. Chronic limb-threatening ischemia and severe claudication behave differently, so a peripheral study either restricts to one or reports them separately.
  • Follow-up. At least 12 months for patency claims, with durability, dilatation and late infection left to PMCF and registries.

Our vascular device CRO team runs the investigations, and our CER team, which includes former Notified Body reviewers, writes and challenges the clinical evaluation.

Check your graft family before the Notified Body does

Send us the product list with positions, coatings and diameters, the IFU and the current CER. We tell you the class of each graft, whether a clinical investigation can be avoided, which benchmarks and endpoints fit each claim, and what PMCF is proportionate.

Request a vascular graft evidence review

05

Questions manufacturers ask

What class is a vascular graft under the EU MDR?

Class III if it is in contact with the aorta or another part of the central circulatory system, or if it contains collagen, gelatin or heparin. An uncoated synthetic graft for a peripheral artery is class IIb under Rule 8.

Does a vascular graft need a clinical investigation under the MDR?

By default, yes, because it is implantable. Article 61 allows exceptions for legacy devices with sufficient clinical data and for equivalence to a device the manufacturer has full access to.

Why are collagen-coated grafts class III?

Collagen and gelatin are derivatives of animal origin, and Rule 18 places devices made with non-viable animal tissues or derivatives in class III.

What are the usual endpoints for a peripheral bypass graft study?

Primary, assisted primary and secondary patency at 12 months, major amputation, amputation-free survival, reintervention, mortality and device failure.

Do vascular grafts need an implant card?

Yes. Article 18 exempts only a short list of implants such as sutures, staples, screws and plates, and vascular grafts are not on it.

Sources and scope

  • Regulation (EU) 2017/745, Articles 18, 54 and 61, Annex VIII (definitions and Rules 8, 14 and 18) and Annex XIV
  • ISO 7198, tubular vascular grafts and vascular patches
  • Global Vascular Guidelines on the management of chronic limb-threatening ischemia, 2019
  • Rutherford et al., recommended standards for reports dealing with lower extremity ischemia, J Vasc Surg 1997

This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The projects described are reported without the manufacturer, the devices or the benchmark products.

Reforming Clinical Evaluation of Medical Devices in Europe