Enteral nutrition · Clinical evaluation · EU MDR
A PEG tube stays in the abdominal wall for months, so the MDR treats it as an implant. Since June 2026, feeding tubes and gastrostomy buttons are on the list of well-established technologies, which removes the default clinical investigation. It does not remove the questions a reviewer asks about a family that has changed connectors, added accessories and been on the market for decades.
Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 6 minute read
Request a gastrostomy CER review Go to the gaps in legacy files01
Under the MDR, a device partly introduced by clinical intervention and left in place for at least 30 days counts as implantable. A PEG tube meets that definition and is class IIb under Rule 8. Article 61(4) made a clinical investigation the default for implantable devices, and Delegated Regulation (EU) 2026/1451 now lists feeding tubes and gastrostomy buttons in Article 61(6)(b), so existing clinical data can carry the file if they are sufficient.
For a family on the market since the 1990s, the data usually exist. What decides the review is whether they still apply to the devices sold today, after connector changes and new accessories.
02
A PEG family is rarely one device. The MDR classifies each device, and each accessory sold separately, in its own right.
| Device | Usual class | What it means for the file |
|---|---|---|
| PEG tube set | Class IIb, Rule 8, implantable | Clinical data on the tube in the indications claimed; now on the WET list as a feeding tube |
| Jejunal extension through an existing PEG | Often class IIb | Its own claims and data, since jejunal feeding carries its own complications |
| Replacement gastrostomy buttons | Usually class IIb | Listed in Article 61(6)(b); data on balloon and non-balloon versions as claimed |
| Clamps, retention discs, connectors sold separately | Class I, or class I sterile | Their own technical file; a sterile class I accessory needs a Notified Body for its sterile aspects |
03
A clinical evaluation plan we prepared for a long-established enteral nutrition manufacturer shows the usual starting point: a PEG family CE marked since the late 1990s, a clinical evaluation built on an old expert report and literature review, and a connector change made since the original certification.
| In the file | The reviewer's question | How to close it |
|---|---|---|
| An expert report from the Directive era | Is the evaluation planned, systematic and up to date as Annex XIV requires? | A new plan and report with a reproducible search, appraisal and state of the art |
| Move from Luer or funnel connectors to ENFit | Do the data collected before the change still apply to the device on sale? | A comparison of the change, bench data on flow and connection, and post-change data from PMS and literature |
| Literature on PEG tubes in general | Which data are on this device, and in which populations? | Device-specific data where they exist, and a justification of how the general literature applies |
| Pediatric use in the IFU | Is there pediatric data for this claim? | Pediatric data, or a narrower IFU |
The ENFit change is worth handling explicitly. ISO 80369-3 small-bore connectors were introduced to prevent misconnections between enteral and other lines, which is a safety benefit the CER can claim, but the reviewer will still want to see that the change did not create new problems with flow, occlusion or dosing through the tube. Our comparison of equivalence and legacy data routes covers how to argue continuity after a design change.
04
The European guidelines on endoscopic management of enteral tubes and on home enteral nutrition shape what a reviewer expects to see benchmarked.
For PMCF, complaint trending, literature monitoring and a structured survey of gastroenterologists and nutrition nurses usually fit a well-established PEG family, with a study kept for claims the data do not cover. See how we design PMCF surveys, PMCF under the MDR, and our guide to the 2026 WET list.
Send us the product list, the IFUs and the current CER. Our clinical evaluation team, which includes former Notified Body reviewers, tells you which devices can rely on the WET route, where design changes need bridging and what PMCF is proportionate.
Request a gastrostomy CER review05
Yes. The MDR treats a device partly introduced by clinical intervention and left in place for at least 30 days as implantable. A PEG tube is usually class IIb under Rule 8.
Yes. Delegated Regulation (EU) 2026/1451 added feeding tubes and gastrostomy buttons to Article 61(6)(b) in June 2026.
No. It removes the default clinical investigation. A clinical evaluation based on sufficient clinical data, a PMCF plan and its updates are still required.
Yes. The CER should show that data collected before the change still apply, with bench data on flow and connection and post-change data from surveillance and the literature.
Each accessory sold separately is classified in its own right. Clamps and retention discs are usually class I, and a sterile class I accessory needs a Notified Body for its sterile aspects.
This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The case described is anonymized and reported at the level of the evidence pattern only.