Enteral nutrition · Clinical evaluation · EU MDR

PEG and gastrostomy tubes under the EU MDR: what the 2026 WET list changes, and what still needs data

A PEG tube stays in the abdominal wall for months, so the MDR treats it as an implant. Since June 2026, feeding tubes and gastrostomy buttons are on the list of well-established technologies, which removes the default clinical investigation. It does not remove the questions a reviewer asks about a family that has changed connectors, added accessories and been on the market for decades.

Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 6 minute read

Request a gastrostomy CER review Go to the gaps in legacy files

01

The short answer

Under the MDR, a device partly introduced by clinical intervention and left in place for at least 30 days counts as implantable. A PEG tube meets that definition and is class IIb under Rule 8. Article 61(4) made a clinical investigation the default for implantable devices, and Delegated Regulation (EU) 2026/1451 now lists feeding tubes and gastrostomy buttons in Article 61(6)(b), so existing clinical data can carry the file if they are sufficient.

For a family on the market since the 1990s, the data usually exist. What decides the review is whether they still apply to the devices sold today, after connector changes and new accessories.

02

Classification, family by family

A PEG family is rarely one device. The MDR classifies each device, and each accessory sold separately, in its own right.

Typical classification of a PEG family under the MDR
DeviceUsual classWhat it means for the file
PEG tube setClass IIb, Rule 8, implantableClinical data on the tube in the indications claimed; now on the WET list as a feeding tube
Jejunal extension through an existing PEGOften class IIbIts own claims and data, since jejunal feeding carries its own complications
Replacement gastrostomy buttonsUsually class IIbListed in Article 61(6)(b); data on balloon and non-balloon versions as claimed
Clamps, retention discs, connectors sold separatelyClass I, or class I sterileTheir own technical file; a sterile class I accessory needs a Notified Body for its sterile aspects

03

The gaps in legacy PEG files

A clinical evaluation plan we prepared for a long-established enteral nutrition manufacturer shows the usual starting point: a PEG family CE marked since the late 1990s, a clinical evaluation built on an old expert report and literature review, and a connector change made since the original certification.

Where a legacy PEG file is usually challenged, and how to close it
In the fileThe reviewer's questionHow to close it
An expert report from the Directive eraIs the evaluation planned, systematic and up to date as Annex XIV requires?A new plan and report with a reproducible search, appraisal and state of the art
Move from Luer or funnel connectors to ENFitDo the data collected before the change still apply to the device on sale?A comparison of the change, bench data on flow and connection, and post-change data from PMS and literature
Literature on PEG tubes in generalWhich data are on this device, and in which populations?Device-specific data where they exist, and a justification of how the general literature applies
Pediatric use in the IFUIs there pediatric data for this claim?Pediatric data, or a narrower IFU

The ENFit change is worth handling explicitly. ISO 80369-3 small-bore connectors were introduced to prevent misconnections between enteral and other lines, which is a safety benefit the CER can claim, but the reviewer will still want to see that the change did not create new problems with flow, occlusion or dosing through the tube. Our comparison of equivalence and legacy data routes covers how to argue continuity after a design change.

04

Endpoints the state of the art sets

The European guidelines on endoscopic management of enteral tubes and on home enteral nutrition shape what a reviewer expects to see benchmarked.

  • Performance. Placement success, time to first feeding, and tube survival before replacement.
  • Procedure-related complications. Peristomal infection, bleeding, perforation and early dislodgement.
  • Device-related complications. Buried bumper syndrome, leakage, occlusion, tube degradation and late dislodgement.
  • Mortality. Reported with care, separating procedure-related deaths from those linked to the underlying disease, which in this population is often severe.

For PMCF, complaint trending, literature monitoring and a structured survey of gastroenterologists and nutrition nurses usually fit a well-established PEG family, with a study kept for claims the data do not cover. See how we design PMCF surveys, PMCF under the MDR, and our guide to the 2026 WET list.

Check your family against the 2026 rules

Send us the product list, the IFUs and the current CER. Our clinical evaluation team, which includes former Notified Body reviewers, tells you which devices can rely on the WET route, where design changes need bridging and what PMCF is proportionate.

Request a gastrostomy CER review

05

Questions manufacturers ask

Is a PEG tube an implantable device under the EU MDR?

Yes. The MDR treats a device partly introduced by clinical intervention and left in place for at least 30 days as implantable. A PEG tube is usually class IIb under Rule 8.

Are feeding tubes and gastrostomy buttons on the MDR WET list?

Yes. Delegated Regulation (EU) 2026/1451 added feeding tubes and gastrostomy buttons to Article 61(6)(b) in June 2026.

Does the WET listing remove the need for a CER on a PEG tube?

No. It removes the default clinical investigation. A clinical evaluation based on sufficient clinical data, a PMCF plan and its updates are still required.

Does a change to ENFit connectors affect the clinical evaluation?

Yes. The CER should show that data collected before the change still apply, with bench data on flow and connection and post-change data from surveillance and the literature.

How are PEG accessories classified?

Each accessory sold separately is classified in its own right. Clamps and retention discs are usually class I, and a sterile class I accessory needs a Notified Body for its sterile aspects.

Sources and scope

  • Regulation (EU) 2017/745, Article 2 (definition of implantable device), Article 61, Annex VIII Rule 8 and Annex XIV
  • Delegated Regulation (EU) 2026/1451 amending Article 61(6)(b)
  • ISO 80369-3, small-bore connectors for enteral applications
  • European Society of Gastrointestinal Endoscopy, guideline on endoscopic management of enteral tubes in adults, 2021

This article is general information for clinical and regulatory teams. It does not predict a conformity assessment outcome. The case described is anonymized and reported at the level of the evidence pattern only.