Guide · Breakthrough devices (BtX) · MDCG 2025-9

EU breakthrough device designation, and what it changes for your clinical evidence

Since December 2025, MDCG 2025-9 gives highly novel medical devices and IVDs a defined route to breakthrough (BtX) status in the EU. Designation brings earlier advice and prioritised review. It does not lower the evidence bar, and it moves part of the evidence burden into post-market follow-up.

MDCG 2025-9Two cumulative criteriaEMA expert panel pilotPMCF after designation
Clinical investigation of a novel medical device in the operating room

Published with TÜV SÜD

Breakthrough device white paper

Dr. Langhit Kurar, TÜV SÜD, and Dr. Nikhil Khadabadi, Eclevar MedTech · September 2026

The ten points

How designation works, point by point.

Each point is drawn from MDCG 2025-9 and from the European Medicines Agency pilot for breakthrough devices. Where the guidance leaves a decision to the manufacturer, we say so.

01 / 10

What BtX designation is

MDCG 2025-9, published by the Medical Device Coordination Group in December 2025, sets out how a medical device or IVD can be recognised as a breakthrough device (BtX) under Regulations (EU) 2017/745 and 2017/746. Its purpose is timely patient access to genuinely novel technologies without compromising safety and performance requirements.

The guidance covers devices and IVDs across technologies and risk classes. Custom-made devices, devices made and used in-house by health institutions, and Annex XVI products without a medical purpose are outside its scope. A device can be both a breakthrough device and an orphan device, in which case MDCG 2025-9 and MDCG 2024-10 apply together.

Regulatory basis

  • MDCG 2025-9, December 2025
  • Regulation (EU) 2017/745 (MDR)
  • Regulation (EU) 2017/746 (IVDR)
  • MDCG 2024-10, orphan devices

02 / 10

Two criteria, both required

A device qualifies only if it meets two cumulative criteria: a high degree of novelty, and a significant positive clinical impact in a life-threatening or irreversibly debilitating disease or condition. Meeting one without the other is not enough.

The request therefore has to show the link between the two: why the novelty is expected to change outcomes for patients, not only how the device differs from what exists.

Regulatory basis

  • MDCG 2025-9, breakthrough device criteria
  • Novelty and clinical impact, cumulative

03 / 10

What counts as novelty

Novelty can sit in the technology, the material, the design, the manufacturing process, the mechanism of action, the biomarker or analyte, the sample type, automation or AI, the clinical procedure, or the way the device is applied in clinical practice. Incremental features and routine iterations of existing devices are unlikely to qualify.

The expert panels' request template asks the manufacturer to position the device on that scale, from a device that differs significantly from available alternatives and the state of the art to a transformative innovation that represents a paradigm shift in a health area. The state of the art analysis in your clinical evaluation is where that position is defended.

Regulatory basis

  • MDCG 2025-9, novelty criterion
  • EMA breakthrough status request template
  • State of the art, MDR Annex XIV

04 / 10

What counts as clinical impact

The device must be reasonably expected to deliver a clinically meaningful improvement for patients with a life-threatening or irreversibly debilitating condition, either compared with the available alternatives and the state of the art, or by addressing an unmet medical need where alternatives are absent or insufficient.

Faster workflow, lower cost, portability or scalability can support the case only where they translate into better outcomes for patients or for public health. A cheaper device that produces the same outcomes is not a breakthrough under this definition.

Regulatory basis

  • MDCG 2025-9, clinical impact criterion
  • Unmet medical need
  • Comparison with available alternatives

05 / 10

How designation is requested

The manufacturer requests an opinion on breakthrough status from the medical device expert panels established under MDR Article 106. A request can be made at any stage of development, once there is enough evidence to address both criteria, and the panels aim to deliver their opinion within 60 days.

A complete clinical data set is not required at this point. Literature, scientific validity, pre-clinical and preliminary clinical or performance data, together with a state of the art analysis, can be enough to make the expected impact credible. The quality of that early evidence is what the panel reads.

Regulatory basis

  • MDR Article 106, expert panels
  • Opinion within 60 days, target
  • Pre-clinical and preliminary clinical data

06 / 10

The EMA expert panel pilot

The European Medicines Agency launched its pilot for breakthrough medical devices on 28 April 2026. It runs in three phases until 2027, subject to available resources, and the advice it provides is free. Phase Ia covered class III devices and class IIb active devices intended to administer or remove a medicinal product, and is closed for status requests. Phase Ib opened on 1 September 2026 for the same device scope, with weekly submission slots until the end of 2026.

Phase II is expected in the first quarter of 2027 and extends requests to all classes of medical devices. Phase III, expected in the third quarter of 2027, extends them to IVDs. For designated devices, the panels can advise on pre-clinical evidence, the clinical development strategy, proposed clinical investigations and post-market clinical follow-up. Notified bodies can submit their own questions, and for class III implantable devices a parallel joint scientific consultation with HTA bodies is possible.

Regulatory basis

  • EMA pilot, launched 28 April 2026
  • Phase Ib from 1 September 2026
  • Phase II expected Q1 2027
  • Phase III, IVDs, expected Q3 2027

07 / 10

What designation changes, and what it does not

Designation brings earlier and more coordinated regulatory interaction: early advice on the clinical development or performance study strategy, scientific advice under MDR Article 61(2) for class III and certain class IIb active devices, coordination with the notified body and, where it applies, the clinical evaluation consultation procedure, and prioritised, structured conformity assessment by the notified body.

It does not mean CE marking, conformity, a guaranteed certificate or market exclusivity. A designation is not automatically lost when a competing device reaches the market, provided the device still meets both criteria.

Regulatory basis

  • MDR Article 61(2), scientific advice
  • Clinical evaluation consultation procedure
  • No market exclusivity

08 / 10

Clinical evidence before CE marking

The pre-market file must still demonstrate safety, performance and clinical benefit. Verification and validation, usability, manufacturing validation, the clinical or performance evaluation and the comparison with the state of the art all remain required.

The clinical evaluation report has to explain the breakthrough case: the advice received, the pre-market evidence, the residual uncertainty and when the evaluation will be updated. For a novel device, that pre-market evidence often includes a first-in-human study or an early feasibility study designed around the questions the panel and the notified body will ask.

Regulatory basis

  • MDR Article 61 and Annex XIV
  • Clinical evaluation report
  • ISO 14155:2026, clinical investigations

09 / 10

Residual uncertainty and PMCF

The guidance accepts that some uncertainty can be moved to the post-market phase through a defined plan, balancing earlier access against the evidence still needed. The notified body may attach conditions to the certificate to manage that residual uncertainty: defined PMCF or PMPF studies, enhanced monitoring, milestone reporting, user information or more frequent review.

Those conditions are binding, and failing to complete them can ultimately affect the validity of the certificate. A breakthrough device therefore needs a PMCF design that is credible before certification, not drafted after it: the endpoints, the follow-up and the data source, whether a prospective study, a registry or both, have to be agreed while the pre-market file is built.

Regulatory basis

  • Certificate conditions
  • PMCF and PMPF
  • MDCG 2020-7 and 2020-8

10 / 10

Preparing the request

A credible request starts with five pieces of work: a precise intended purpose and target condition, a state of the art analysis that shows where alternatives fall short, a clear statement of where the novelty sits, an estimate of the expected clinical impact against those alternatives, and an evidence plan that shows what will be known before and after CE marking.

Engage the notified body early, check the pilot's current eligibility and submission slots with EMA before setting dates, and treat the panel's advice as an input to your regulatory and clinical strategy rather than a separate exercise. Eclevar's team includes former notified body clinical reviewers who build that plan with you, from the first clinical investigation to PMCF.

Regulatory basis

  • Intended purpose and target condition
  • State of the art
  • Evidence plan, pre and post market

Sources: MDCG 2025-9, Guidance on Breakthrough Devices (BtX), December 2025; EMA, expert panel support for breakthrough medical devices: pilot programme. Pilot dates as published by EMA in November 2026.

Review of clinical data for a novel medical device
Designation is not certificationThe evidence bar stays the same. What changes is when, and with whom, you agree how to clear it.

Talk to a former reviewer

Test your breakthrough case before you file.

Bring your intended purpose, your state of the art and your current evidence. We will show you where the novelty and impact case is strong, where it is thin, and what the evidence plan needs.

Book a scoping call

Published with TÜV SÜD

A notified body and a CRO on the same pathway.

Eclevar co-authored TÜV SÜD's white paper on breakthrough devices under the EU MDR: Dr. Nikhil Khadabadi, Eclevar's Chief Medical Officer at the time of publication, wrote it with a TÜV SÜD clinical reviewer. It is published on TÜV SÜD's knowledge hub.

Medical Breakthrough Devices (BtX)

Supporting earlier market access for novel medical devices. White paper, TÜV SÜD, 17 September 2026.

Read it on tuvsud.com →

Dr. Langhit Kurar

Clinical Reviewer and Orthopaedic Subject Matter Expert, TÜV SÜD.

Dr. Nikhil Khadabadi

Eclevar MedTech, Chief Medical Officer at publication. Orthopaedic surgeon and former notified body clinical reviewer.

Co-authorship of a publication does not imply endorsement of Eclevar services by any notified body and has no bearing on any conformity assessment.

Designation, against certification

What BtX status gives you, and what it does not.

Not part of designation

What you still have to earn.

×
CE marking or a certificate.Designation is an opinion on status. Conformity assessment still decides the certificate.
×
A lower evidence standard.Safety, performance and clinical benefit must still be shown before certification.
×
Market exclusivity.Competing devices can also be designated and certified.
×
Freedom from post-market conditions.Residual uncertainty returns as binding PMCF and reporting conditions.

What designation brings

Earlier, structured dialogue.

✓
An expert panel opinion on status.Targeted within 60 days of the request.
✓
Advice on the evidence strategy.Pre-clinical evidence, clinical development, investigations and PMCF.
✓
Prioritised notified body review.Structured, prioritised conformity assessment work.
✓
A defined place for uncertainty.A plan that stages part of the evidence after market access.

Plan your breakthrough evidence with former notified body reviewers.

Bring your device and your current evidence. We map the request, the clinical investigations and the PMCF that follows.

Book a free scoping call

Questions sponsors ask us first

Breakthrough designation, the practical questions.

Is EU breakthrough designation the same as the FDA Breakthrough Devices Program?
No. They are separate programs with separate criteria and procedures. In the EU, the opinion on status comes from the medical device expert panels under MDCG 2025-9. In the US, the FDA decides on designation under its own program. Neither is a marketing authorization: an EU breakthrough device still needs a CE certificate, and a US one still needs FDA marketing authorization.
Does designation lower the clinical evidence required for CE marking?
No. The device must still show safety, performance and clinical benefit before certification. What changes is that some residual uncertainty can be addressed after market access through a defined plan, usually backed by binding post-market conditions on the certificate.
Which devices can apply now?
In the EMA pilot, phase Ib opened on 1 September 2026 for class III devices and class IIb active devices intended to administer or remove a medicinal product. Phase II, expected in the first quarter of 2027, extends requests to all device classes, and phase III, expected in the third quarter of 2027, to IVDs. Check current submission slots with EMA before planning dates.
Can a device be both a breakthrough device and an orphan device?
Yes. MDCG 2025-9 confirms that it and MDCG 2024-10 on orphan devices can apply in parallel, so eligibility for both should be assessed together for small-population, high-need indications.
What happens if the post-market conditions are not met?
The conditions attached to the certificate are binding. Failing to complete them, for example a defined PMCF study, can ultimately affect the validity of the certificate. That is why the PMCF design should be agreed before certification.
How does Eclevar help with a breakthrough device?
We build the evidence plan behind the request: the state of the art analysis, the first-in-human or early feasibility study, the clinical evaluation report and the PMCF that certificate conditions may require. Our team includes former notified body clinical reviewers. Designation and certification decisions remain with the expert panels and the notified body.

Reforming Clinical Evaluation of Medical Devices in Europe