Spine · Clinical evaluation · EU MDR
Sham-controlled trials have questioned whether vertebroplasty helps more than a placebo procedure. Under the MDR, that debate is part of the state of the art every cement manufacturer has to answer, and case series showing that pain went down no longer close the question.
Eclevar MedTech clinical evaluation team · Updated November 26, 2026 · 7 minute read
Request a cement CER gap review Go to the gap in cement files01
A cement for vertebral augmentation is an implant that stays for life, and its clinical evaluation has to show a clinical benefit measured against the state of the art. For osteoporotic fractures, that state of the art includes a Cochrane review that found no clinically important benefit over a sham procedure in routine practice. A CER built on uncontrolled series of patients whose pain improved cannot separate the effect of the cement from the natural recovery of the fracture.
The bar is also higher than under the Directive. Rule 8 of Annex VIII places implantable devices that come into contact with the spinal column in class III, and MDCG 2021-24 reads that contact as intended contact with any bony structure of the column. A cement injected into a vertebral body is in that position, so expect class III and what follows from it: a clinical investigation by default under Article 61(4), the clinical evaluation consultation procedure, a summary of safety and clinical performance, and for legacy devices an MDR certificate by December 31, 2027.
The files that pass define the benefit for a precise population, set acceptance criteria from controlled data, report safety with clear definitions, and carry a PMCF plan that answers what the literature leaves open.
02
Two sham-controlled trials published in 2009 found that vertebroplasty did no better than a placebo procedure. The Cochrane review updated in 2018 pooled the randomized evidence and concluded, on high- to moderate-quality evidence, that vertebroplasty has no important benefit on pain, disability or quality of life for acute or subacute osteoporotic fractures in routine practice, with no difference by pain duration. It was uncertain whether the procedure increases the risk of new vertebral fractures.
The picture is not uniform. One sham-controlled trial in patients with recent fractures and severe pain reported a benefit, and another in a similar population did not. Pathological fractures from cancer and kyphoplasty with a balloon rest on different bodies of evidence. For a reviewer, this means one thing: the benefit claim has to name the population, the fracture type and the comparator, and the clinical data have to match that claim.
03
A clinical evaluation we prepared for a European maker of PMMA cement shows the usual pattern. The literature on the device was mostly uncontrolled, it covered indications beyond the spine, such as filling bone cysts, and the risk file relied on published measurements of cement temperature during setting. Each of these is a fair source. None of them answers the benefit question on its own.
| In the file | The reviewer's question | How to close it |
|---|---|---|
| Pain reduction from uncontrolled series | How much of the improvement is the cement, and how much natural recovery? | Acceptance criteria taken from the sham and conservative arms of randomized trials, and a comparator in PMCF |
| Literature across several bone indications | Is each indication in the IFU supported by its own data? | Analyze by indication, or narrow the IFU to what the data cover |
| Cement leakage reported in different ways | Is leakage defined, imaged the same way and split into symptomatic and asymptomatic? | One definition, imaging at fixed time points, and pulmonary cement embolism reported separately |
| Setting temperature from published studies | Does the evidence cover this cement's formulation and volumes? | Bench data on the device, linked to the risk file and to any clinical signal of thermal injury |
| Equivalence to another cement | Do composition, radiopacifier, viscosity and setting time match? | A characteristic-by-characteristic comparison, or a claim built on the device's own data |
Viscosity is named in Annex XIV of the MDR as a technical characteristic for equivalence, and for cements it drives leakage. Our comparison of equivalence and legacy data routes sets out the tests.
04
A single-arm PMCF study can work if its performance goals come from controlled data and its population matches the claim. Where the claim is comparative, a comparator arm with conservative care is the stronger design, and our biostatistics team sizes it on the primary endpoint. See how we run PMCF studies under the MDR, and why market history alone rarely carries an orthopedic CER.
05
In the United States, PMMA cements for vertebroplasty are class II under 21 CFR 888.3027 and are cleared through 510(k), in some cases on bench data alone. The FDA has also published clinical trial considerations for vertebral augmentation devices. A cement cleared in the US on bench testing still needs a clinical evaluation under the MDR, so a manufacturer entering Europe should plan the clinical data once, around the endpoints both reviewers read.
Our orthopedics and spine team designs these studies and writes the clinical evaluation report that draws on them.
Send us the CER, the IFU and the PMCF plan. Our clinical evaluation team, which includes former Notified Body reviewers, checks each claim against the state of the art and tells you which can stand on existing data and which need PMCF.
Request a cement CER gap review06
Yes. It is part of the state of the art for osteoporotic fractures. A CER that claims a benefit for that population has to address it and explain why the device's data support the claim.
Not on their own. Pain after an osteoporotic fracture improves over time without treatment, so series without a comparator cannot isolate the effect of the cement. They remain useful for safety and for performance goals drawn from controlled data.
With one definition, imaging at fixed time points, and a split between asymptomatic leakage, symptomatic leakage and pulmonary cement embolism, compared with rates from the literature.
Only if each indication in the IFU is supported by its own clinical data. Data from bone cysts or other defects do not support a vertebral fracture claim, and the reverse is also true.
No. A 510(k) can rest on bench data. The MDR requires a clinical evaluation based on sufficient clinical data for every claim, so US clearance is a starting point for the European file, not a substitute.
This article is general information for clinical and regulatory teams. It makes no claim about the treatment of any patient and does not predict a conformity assessment outcome. The case described is anonymized and reported at the level of the evidence pattern only.