Do we need clinical data for every procedure on the label?
Each procedure needs clinical data or a justified extension from related procedures. Notified bodies look closely at indications that go beyond the procedures actually studied.
Can we claim equivalence with another surgical robot?
Rarely in practice. Equivalence requires comparable technical, biological and clinical characteristics, and differences in software, kinematics and instruments usually prevent it. Most robotics CERs rely on data on the system itself.
How should the learning curve appear in the CER?
As evidence, not as a caveat. The CER shows how outcomes change with experience, what training and proctoring are required, and how the PMCF plan follows results for new users.
Does the EU AI Act apply to our robot?
It can, when the system includes AI functions that qualify as high-risk under the AI Act. Each function is assessed individually, and the AI Act documentation and the clinical evaluation have to describe the same functions and performance.
Can U.S. clinical data support the EU CER?
Yes, when the system version, procedures, surgeon training and patient population are applicable to the EU claims. The CER describes the differences and their effect.
Who stays responsible for the clinical evaluation?
The manufacturer. ECLEVAR plans, writes and challenges the documents, and supports you during the review. The conclusions on benefit-risk and conformity remain yours.