Does a wound dressing need a clinical investigation under the MDR?
It depends on the class and on the data available. For Class III dressings, such as those with an ancillary medicinal substance or made from animal tissue, the MDR expects a clinical investigation unless one of its exemptions applies, for example equivalence under the conditions of Article 61(5). For Class I, IIa and IIb dressings, the clinical evaluation can rest on literature, equivalence and post-market data if together they are sufficient to support the claims.
Can we claim equivalence with another manufacturer's dressing?
Equivalence has to be shown on technical, biological and clinical characteristics, as described in MDCG 2020-5. For Class III devices, equivalence with another manufacturer's device also requires a contract giving ongoing full access to its technical documentation. In practice, equivalence with a competitor's dressing is often hard to support, and the gap review tells you early whether it is realistic.
Is our MDD-era CER still usable?
As a starting point. Legacy files usually need data on the device itself, an updated state of the art, a check that the claims still match the evidence, and a PMCF plan that targets what remains open. MDCG 2020-6 describes what notified bodies expect for legacy devices.
How is a silver or antimicrobial dressing classified?
It depends on what the antimicrobial does. If it is a medicinal substance acting on the wound with an action ancillary to the dressing, Rule 14 places the device in Class III and a medicines authority is consulted on the substance. The rationale is settled with your notified body before the CER is written, because the class sets the evidence the CER must carry.
Can U.S. real-world data support a European dressing CER?
It can contribute when the records identify your dressing, the population matches the intended purpose and the methods are fixed before analysis. The applicability of U.S. data has to be justified in the clinical evaluation. Whether the evidence is sufficient is the notified body's decision.
Who stays responsible for the clinical evaluation?
The manufacturer. ECLEVAR plans, writes and challenges the documents, and supports you during the review. The conclusions on benefit-risk and conformity remain yours.