Is compliance with ISO 80601 standards enough for a CER?
No. Device standards establish bench performance and basic safety. The clinical evaluation has to show clinical performance, safety and benefit in the intended patients and settings, using clinical data on the device or an equivalent one.
Does our ventilator CER need a clinical investigation?
Not necessarily. Class IIb devices can rely on literature, equivalence and post-market data when together they are sufficient. A clinical investigation becomes likely when the device introduces a new ventilation mode, a new population or a claim that existing data cannot support.
Does the expert panel consultation apply to our device?
It can apply to Class IIb active devices that administer or remove a medicinal product, unless an exemption applies. Whether it applies depends on the intended purpose and the substances involved, and it is settled with your notified body early.
How do we use years of post-market experience?
Post-market data can be strong evidence when it is organized: exposure estimated, complaints and incidents analyzed by type and severity, and results compared with the state of the art. Raw complaint counts without exposure or analysis carry little weight.
Can home-use data come from outside the EU?
Yes, when the population, setting and care pathway are shown to be applicable. Home care practice differs between countries, so the CER describes those differences and their effect.
Who stays responsible for the clinical evaluation?
The manufacturer. ECLEVAR plans, writes and challenges the documents, and supports you during the review. The conclusions on benefit-risk and conformity remain yours.