Does our software need a clinical investigation?
Not always. Many software devices close the clinical performance pillar with retrospective validation on well-characterized data. Class III software is expected to have a clinical investigation unless an MDR exemption applies, and a prospective study may be needed at lower classes when retrospective data cannot show performance in the intended setting.
Is technical validation enough for a CER?
No. MDCG 2020-1 expects evidence of valid clinical association, technical performance and clinical performance. Technical validation shows that the software computes its output correctly; it does not show that the output is clinically meaningful in the intended population.
How do we handle new versions of the software?
Each change is assessed for its effect on performance and on the intended purpose. Some changes need repeated validation and an updated CER, and significant changes can require notified body involvement before release. The PMCF plan should set out how versions are monitored.
Does the EU AI Act change the clinical evaluation?
The AI Act adds its own requirements for AI-based medical device software, including data governance, transparency and human oversight. It does not replace the MDR clinical evaluation. Both sets of documents have to describe the same intended purpose, data and performance.
Can the clinical evaluation rely on data from outside the EU?
Yes, when the population, setting of use, input devices and clinical practice are shown to be applicable to the intended use in the EU. Differences are described in the CER and, where they matter, tested in subgroup analyses.
Who stays responsible for the clinical evaluation?
The manufacturer. ECLEVAR plans, writes and challenges the documents, and supports you during the review. The conclusions on benefit-risk and conformity remain yours.