Wound real-world evidence

U.S. wound data for CER and PMCF under the EU MDR

A notified body question about clinical benefit or current use does not always need a new clinical investigation. Retrospective U.S. wound records can supply clinical data for a clinical evaluation report or a PMCF plan, when the device is identified and the applicability of the data is justified.

Three questions a reviewer will ask

Is it your device?

Data on the device itself, or on a device for which equivalence is demonstrated.

Is it your population?

Patients and conditions of use that match the intended purpose.

Was it planned?

Objectives, endpoints and analysis fixed before the data were examined.

The feasibility review answers all three before any study is commissioned.

Work aligned withEU MDR 2017/745ISO 14155:2026ISO 1348521 CFR Part 11GDPR

Where it fits

Three uses in a clinical evaluation

Under the EU MDR, clinical evaluation is a continuous process, updated with post-market data. Retrospective records can feed it at three points.

Closing a gap in the CER

A targeted analysis that addresses a specific question on clinical benefit, performance or safety in an indication where current evidence is thin.

PMCF activity

A retrospective study defined in the PMCF plan as a specific method, with its rationale, objectives and analysis described in advance.

Response to a deficiency

Evidence prepared for a specific question raised during conformity assessment, with a study report the reviewer can trace back to the source data.

Applicability

Making U.S. data count in a European file

Data from outside the EU can be used in a clinical evaluation. The manufacturer has to show that they apply. These are the points we document.

Device identification

The records identify your device, by brand and where possible by model. Category-level records support a description of the state of the art, not a claim about your device.

Same device, same version

Any difference between the product used in the United States and the CE-marked device is assessed and documented.

Population and intended purpose

Indication, wound type and patient characteristics are compared with the intended purpose in your technical documentation.

Clinical practice

Differences in standard care between the United States and Europe, such as care setting and product availability, are described and their effect discussed.

Data quality and governance

Provenance, completeness, permitted uses and data protection are documented so that the reviewer can assess reliability.

Outputs

What goes into which document

The study feeds the regulatory documents. It does not replace the manufacturer's own conclusions.

DocumentWhat the study contributesWhat stays with the manufacturer
Clinical evaluation reportClinical data on performance, benefit and safety in routine use, with an appraisal of quality and relevance.The overall benefit-risk conclusion and the demonstration of conformity.
PMCF planThe method, rationale, objectives, endpoints and timelines of the retrospective study.The link to residual risks and open questions identified in the clinical evaluation.
PMCF evaluation reportResults, limitations and conclusions of the study.Decisions on updates to the CER, risk management and labeling.
Deficiency responseA study report and an evidence memo addressing the question raised.The formal response and the dialogue with the notified body.

Pitfalls

Why reviewers set retrospective data aside

Retrospective data are often submitted and often discounted. These are the usual reasons, and each is checked during feasibility.

The device is not identified

Data pooled across a product category say little about one manufacturer's device.

No pre-specified plan

An analysis designed after the data were seen reads as a search for a favorable result.

Applicability asserted, not shown

The file states that U.S. data apply without comparing populations and practice.

Missing data left unexplained

Reviewers expect to see how much is missing and what it does to the conclusions.

Safety not addressed

Routine wound records capture complications unevenly. The report states what can and cannot be concluded on safety.

Conclusions beyond the data

Claims broader than the cohort, the indication or the follow-up actually studied.

Track record

The numbers behind the method

Of the 56 studies ECLEVAR has delivered, 21 are PMCF studies and registries. The methods on this page come from that work.

56

Studies delivered

As of November 2026

By study phase

Studies by country

France18
Germany12
United Kingdom10
Italy8
Spain6
Nordic market3
Ireland2
Other5

A study running in several countries is counted once in each country.

400+

Devices assessed by former notified body reviewers now in ECLEVAR leadership

9

Clinical monitors (CRAs)

4

Project managers

3

Medical writers

3

Data management and biostatistics

2

Quality assurance

Clients

Medical device manufacturers we work with

Manufacturers of wound care products, implants, monitoring systems and dental devices use ECLEVAR for clinical investigations, PMCF and registries. More in our client success stories.

Client: RegenLab

Multicountry prospective PMCF in wound care

A two-cohort program in chronic wounds, diabetic foot ulcers and venous leg ulcers, run across European sites under one protocol and one data model.

160

participants

14

sites

5

European countries

Client: Perouse Medical, Vygon Group

Six registered observational studies across a vascular portfolio

ECLEVAR structured the program architecture and supports delivery. Intended evidence use: clinical evaluation and PMCF.

6

registered observational studies

3

device families

1,150+

subjects

5 years

follow-up

Team

The people you will speak to

A small senior team, with U.S. clinical consultants working alongside European delivery for sponsors who plan evidence on both sides of the Atlantic.

Portrait of Michael McGuffey

Michael McGuffey

U.S. Clinical Consultant

Former clinical lead for final labeling on Medtronic's Hugo robotic-assisted surgery launch team, where he helped design surgeon training and proctoring and started a clinical registry for the platform.

Portrait of Dawn Heimer

Dawn Heimer, PhD

U.S. Clinical and Regulatory Consultant

Provides strategic input on U.S. medical-device clinical operations, regulatory considerations and real-world evidence programs.

Portrait of Chems Hachani

Chems Hachani

Founder and CEO

Sets the evidence strategy with the sponsor and stays accountable for delivery from first scoping call to final report.

Portrait of Karina Schönborn

Karina Schönborn

Business Development Lead

Your first point of contact for scoping, budget scenarios and proposals.

ECLEVAR leadership includes former notified body reviewers; the full team is on the leadership page. Platinum Award · CVBF & MILO Health, xShare / EUCROF EHDS & Clinical Research Open Call. The award recognizes the submitted data use case and is not an evaluation of CRO services.

Questions

What sponsors ask before they start

Can data collected in the United States be used in a clinical evaluation under the EU MDR?

Yes, provided the manufacturer justifies that the data apply to the device, the population and the conditions of use covered by the CE marking. The clinical evaluation has to appraise the quality and relevance of the data. Whether the evidence is sufficient is the notified body's decision.

Does a retrospective study count as PMCF?

It can. Annex XIV Part B of Regulation (EU) 2017/745 expects the PMCF plan to set out the general and specific methods to be used, with their rationale. A retrospective study of real-world records can be one of those specific methods when it addresses a question identified in the clinical evaluation.

Our notified body raised a deficiency. Is a retrospective study fast enough?

It depends on the question and on the data. The target is eight weeks from kickoff once feasibility confirms that the records can answer the question within a fixed scope. The feasibility review comes first, and it can conclude that another route is needed.

Who writes the CER?

ECLEVAR can deliver the study report and the CER input, or update the clinical evaluation report itself as a separate scope. The manufacturer remains responsible for the clinical evaluation and its conclusions.

Next step

Discuss your wound care product

Send the product name, wound indication, outcome and follow-up period you have in mind. If you need a comparison, tell us what clinicians typically use instead.

We aim to respond within one business day. A confidentiality agreement can be signed before you share any document.

Share the decision question

Deficiency letter, PMCF or CER gap, reimbursement question or comparator need.

Open the data-fit discussion

Under a confidentiality agreement, we review the data source, variables, completeness and cohort logic.

Lock the route and scope

Go or no-go, evidence model, deliverables, assumptions and fixed-scope kickoff.